Article |

Comparison of Auscultatory and Oscillometric Blood Pressures FREE

Myung K. Park, MD; Shirley W. Menard, RN, PhD; Cheng Yuan, PhD
[+] Author Affiliations

From the Department of Pediatrics, School of Medicine (Dr Park), School of Nursing (Dr Menard), and the Department of Computing Resources (Dr Yuan), University of Texas Health Science Center, San Antonio.


Arch Pediatr Adolesc Med. 2001;155(1):50-53. doi:10.1001/archpedi.155.1.50.
Text Size: A A A
Published online

Objective  To study the differences in blood pressure readings between the auscultatory and oscillometric (Dinamap model 8100; Critikon, Tampa, Fla) methods.

Design  Survey of 2 blood pressure instruments.

Setting  Public schools.

Participants  Seven thousand two hundred eight schoolchildren aged 5 through 17 years.

Main Outcome Measure  Blood pressure levels.

Results  For all children combined, Dinamap systolic pressure readings were 10 mm Hg higher (95% confidence interval, −4 to 24 mm Hg) than the auscultatory systolic pressure readings. Dinamap diastolic pressure readings were 5 mm Hg higher (95% confidence interval, −14 to 23 mm Hg) than the auscultatory Korotkoff phase V diastolic pressure readings.

Conclusion  These findings preclude the interchange of readings by the 2 methods. Caution must be exercised in the diagnosis of hypertension when an automated device is used.

Figures in this Article

AN IMPORTANT element in pediatric practice is the routine measurement of blood pressure (BP). Since the accuracy of an oscillometric device in reflecting direct arterial pressure was reported in the neonate,1,2 infant and child,1,3,4 and adult,5 the Dinamap monitor (Critikon, Tampa, Fla) has become widely used in pediatric patient care. The comparison of BP values obtained during office visits against normative values is used as a screen for asymptomatic hypertension, assuming that those values are comparable whether or not the same methods for measurement were used. In a previous study, we reported that the systolic pressure levels obtained by the Dinamap method (model 1846SX) were 6 mm Hg higher (P<.05) than those obtained by the auscultatory method in fifth-grade children.6 It is not known if there is also a clinically important magnitude of difference in BP levels measured by the 2 devices for other age groups or how the newer model (8100) of Dinamap device compares with auscultatory BP measurement. This information is important in view of the wide clinical use of the new model of the Dinamap monitor.

The San Antonio (Tex) Triethnic Children's Blood Pressure Study was conducted to test for possible ethnic differences in BP levels among 3 major ethnic groups of children. We used both the auscultatory method and Dinamap model 8100 because the latter was devoid of observer-related variability in BP readings. We found that there were significant differences between BP readings obtained by the 2 methods. This report addresses the magnitude and clinical implications of these differences.

In the San Antonio Triethnic Children's Blood Pressure Study, both oscillometric and auscultatory BP readings were obtained on 7208 schoolchildren aged 5 through 17 years. A letter of invitation was sent to the parents of each child enrolled in kindergarten through 12th grade. The parents of the participating children signed their consent and the children assented to participate in the study. The University of Texas Health Science Center's institutional review board approved the proposal. There were 3356 boys (46.6%) and 3852 girls (53.4%) of whom 4215 were Mexican American (58.5%), 2040 were non-Hispanic white (28.3%), and 953 were African American (13.2%).

The Dinamap model 8100 was used to measure BP, which was either preceded or followed by auscultatory measurement of BP. The order of the measurement was randomized so that about half of the children had their BP measured by the auscultatory method first followed by the Dinamap method (A rotation), and the remainder of the children had their BP measured first by the Dinamap followed by the auscultatory methods (D rotation). Two Dinamap monitors were used in this study. Research nurses who took oscillometric BP readings were taught to use the Dinamap monitor through in-service training, familiarization with the manual, and "hands-on" practice. The oscillometric device was calibrated for accuracy by the bioengineering department of our institution once every month or whenever a malfunction was suspected.

The auscultatory method of BP measurement used a Baumanometer mercury gravity sphygmomanometer (WA Bauman Co, Copiague, NY). Because of well-known observer-related variability,7 only those research nurses who successfully completed the 6-hour Instructor's Course for Blood Pressure Determination offered by the American Heart Association, Texas Affiliate,8 were selected to measure auscultatory BP. To ensure acceptable consistency among these nurses, a small group of children had their BP measured by several nurses. Two nurses, whose obtained readings were most similar and not statistically different, were selected to do the BP measurements. In the pilot study, the means of the systolic BP readings differed by 0.5 mm Hg (95% confidence interval [CI], –11.2 to 12.2) and the means (95% CI) of the diastolic BP readings differed by 2.4 mm Hg (95% CI, −11.6 to 16.5 mm Hg).

Three BP readings were obtained by each method on the right arm, with the patient sitting. The width of the BP cuff was selected to be 40% to 50% of the circumference of the upper arm as recommended by the American Heart Association9 and the Working Group of the National High Blood Pressure Education Program (NHBPEP).10 This cuff selection method was shown to be appropriate for the oscillometric BP measurement in children by a previous study.4 Blood pressure cuffs manufactured by the Critikon Company were easily adaptable to the Baumanometer and were used for both the auscultatory and oscillometric methods. Dinamap BP readings were unknown to the observers at the auscultatory BP station.

The mean of the 3 BP readings was used for statistical analyses using the SAS Software (SAS Institute Inc, Cary, NC). For the analysis of the difference between BP readings by the auscultatory and the Dinamap methods, the mean of the 3 BP readings by the auscultatory method was subtracted from the mean of those by the Dinamap method for each subject, and the mean and 95% CI of the difference were computed. For diastolic pressure, only the Korotkoff phase V diastolic pressure (K5) was compared with the oscillometric BP readings, as recommended by the Working Group of the NHBPEP.10

Because of an unexpectedly large difference observed in BP readings between the Dinamap model 8100 and auscultatory methods, we compared measurements by this model with those by the earlier model (1846SX) in 2 different ways. First, we compared BP readings obtained by the auscultatory and Dinamap methods in an earlier study6 with those from the current study for fifth-grade children from the same school district. In that study, BP readings obtained by the model 1846SX and the auscultatory method were compared in the fifth-grade children.6 Second, BP readings were obtained by both models of the Dinamap monitor in 48 17-year-old students and the results were compared to test whether the 2 models gave significantly different readings. Triplicate BP readings were obtained in both arms simultaneously using model 1846SX on 1 arm and model 8100 on the other arm. The arm used for each model was alternated from subject to subject.

The difference in BP readings between the 2 different devices was related to the mean of those pressures suggested by Bland and Altman.11 The rating system for automated BP devices recommended by the British Hypertension Society12 was used to evaluate the Dinamap model 8100. This system grades the degree of agreement with auscultatory measurements based on the percentage of comparisons that fall within particular ranges of difference between the 2 measures. A grade of A indicates the differences to be relatively small in most comparisons, whereas B and C grades indicate a progressively lower degree of agreement.12 A grade of D indicates an even lower degree of agreement, (less than 45% of the measurement difference is less than 5 mm Hg12).

Blood pressure readings obtained by the 2 Dinamap monitors (both model 8100) were similar. The mean (95% CI) of the systolic pressure by machine A was 111.4 mm Hg (92.2-130.6 mm Hg) and that by machine B was 112.0 mm Hg (92.8-131.2 mm Hg). Auscultatory BP readings obtained by the 2 observers were also similar. The mean (95% CI) of the systolic pressure obtained by observer 1 was 100.2 mm Hg (80.8-119.6 mm Hg) and that obtained by observer 2 was 102.8 mm Hg (83.0-122.6 mm Hg). The diastolic K5 readings obtained by the 2 observers were 56.2 mm Hg (37.2-75.2 mm Hg) and 56.3 mm Hg (40.6-72.0 mm Hg), respectively. For a given method, BP readings were similar between children who had their BP measured first by the auscultatory method followed by the Dinamap monitor (A rotation) and those children who had their BP measured in the reverse order (D rotation) (Table 1). Regardless of the order of measurements, Dinamap systolic and diastolic pressures were higher than auscultatory measurements for both the A and D rotation groups.

Table Graphic Jump LocationTable 1. Auscultatory and Dinamap Blood Pressure Readings According to the Order of Measurements*

There were large and clinically important differences in BP readings between the Dinamap and auscultatory methods throughout the ages studied (Table 2). The overall average systolic pressure obtained by the Dinamap was consistently higher than that obtained by the auscultatory method. The mean (95% CI) differences were 10.2 mm Hg (–3.5 to 24.0 mm Hg) for systolic pressure and 4.7 mm Hg (–13.5 to 22.9 mm Hg) for diastolic pressure for all age groups combined (Table 2). The mean differences between the 2 BP methods were also similar among the ethnic groups and between the genders (Table 3).

Table Graphic Jump LocationTable 2. Mean Difference Between Dinamap* and Auscultatory Blood Pressure Readings According to Age Group†
Table Graphic Jump LocationTable 3. Mean Difference in Dinamap* and Auscultatory Blood Pressure Readings by Ethnicity and Sex†

Figure 1 shows the difference in systolic pressure between the 2 methods (Dinamap minus auscultatory) in 9- to 10-year-old boys plotted against the means of systolic pressures by the 2 methods. The figure was a typical plot for boys and girls of all age groups and is shown as a representative plot. The figure shows that the difference is independent of the BP readings observed in the study population. The differences in the diastolic pressure were also unrelated to the level of BP.

Place holder to copy figure label and caption
Figure 1.

Magnitude of difference in systolic blood pressure readings by the Dinamap and auscultatory methods plotted against the means of systolic blood pressure readings by the 2 methods for boys aged 9 to 10 years. Two broken horizontal lines represent the 95% confidence intervals for the mean difference. The solid horizontal line represents the mean difference.

Graphic Jump Location

When the results of the current study were compared with those of the earlier study,6 the mean ages of the fifth graders were comparable between the 2 studies (11.4 vs 11.6 years). The group means of systolic and diastolic pressures obtained by the Dinamap model 8100 were 3.5 mm Hg higher and 6.8 mm Hg higher, respectively, than those obtained by the model 1846SX in the 1990 study.6 Auscultatory pressures did not differ significantly between the 2 cohorts. In the current study, the mean auscultatory systolic pressure was 1.2 mm Hg lower and the mean K5 pressure was 1.7 mm Hg lower than those found in the earlier study (Figure 2). When readings obtained by the 2 models of the Dinamap monitor were compared for the 17-year-old subjects, the mean of the 3 BP readings by model 8100 was 1.8 mm Hg higher (95% CI, –10.2 to 13.0) for systolic pressure and 1.3 mm Hg lower (95% CI, –11.4 to 8.8) for diastolic pressure than those obtained by model 1846SX.

Place holder to copy figure label and caption
Figure 2.

Comparison of blood pressure levels of fifth-grade students from 2 studies. Dinamap model 1846SX was used in the 1990 study6 and Dinamap model 8100 (Critikon, Tampa, Fla) in the current study. Although the populations were different, they are both fifth-grade students from the same school district. BP indicates blood pressure; DIN, Dinamap; AUS, auscultatory; and K5, Korotkoff phase V diastolic pressure.

Graphic Jump Location

According to the British Hypertension Society grading system,12 the systolic pressure readings of the Dinamap monitor 8100 achieved grade D. When compared with auscultatory K5 pressures, Dinamap diastolic pressure also achieved grade D. The grade was D each time, whether examined according to sex, ethnicity, or age.

In this study, we demonstrated that there are large, clinically important differences between BP levels measured by the auscultatory method and those measured by the Dinamap model 8100, a popular oscillometric device. Thus, BP obtained by 1 method is not interchangeable with that obtained by the other. Although a significant, but small, difference in BP readings between the 2 methods might have been expected based on our earlier study in fifth-grade children,6 the large difference we observed was unexpected. In the earlier study, using the Dinamap monitor model 1846SX, the mean systolic pressure was 6.4 mm Hg higher than the auscultatory systolic pressure and the mean diastolic pressure was 3.4 mm Hg lower than the auscultatory K5 diastolic pressures.6 In the current study, the magnitude of the difference in systolic pressure was 10 mm Hg, which was higher than that reported in the earlier study,6 whereas the Dinamap diastolic pressure was 5 mm Hg higher than the auscultatory K5 diastolic pressure. The reasons for the large difference in systolic BP levels measured by the 2 methods are not clear. The difference may be due either to higher BP readings by model 8100 than the earlier model 1846SX, lower auscultatory systolic pressure readings by the current research team than the previous team, or a combination of the 2. Different models use different algorithms, measure different quantities,13,14 and thus may result in different BP readings.

O'Brien et al15 compared the Dinamap model 8100 with auscultatory BP readings using BP data from 86 subjects aged 15 to 80 years with a wide range of BP readings. According to the British Hypertension Society evaluation system,12 the Dinamap model 8100 achieved a grade B for systolic pressure and a grade D for diastolic pressure.15 In our study, the Dinamap model 8100 achieved a grade D for both systolic and diastolic pressures. The reason(s) for the different ratings between the study of O'Brien et al and this study are not clear. The different age groups with different BP ranges and different auscultatory BP observers could have been contributing factors. In both children and adults, however, BP readings by the auscultatory and Dinamap methods were shown to be not interchangeable, at least for the model 8100. However, these data do not address whether the auscultatory or oscillometric measurement of BP, or which model of Dinamap, is more accurate. The important point, however, is that different methods can yield different values and that abnormal BP readings must be interpreted with this inconsistency in mind.

In conclusion, systolic pressure levels measured by the Dinamap model 8100 are approximately 10 mm Hg higher than those assessed by the auscultatory method in children aged 5 to 17 years. Dinamap diastolic pressures were on the average 5 mm Hg higher than auscultatory K5 pressures. This finding precludes the interchange of BP readings by the 2 methods and indicates that caution must be exercised in the diagnosis of hypertension when this device is used. Similarly, the BP readings obtained by other automated BP measuring devices should not be considered interchangeable with auscultatory pressure readings.

Accepted for publication August 29, 2000.

This study was supported by grant MCJ-480612-0 from the Maternal and Child Health Program (Title V, Social Security Act), Health Resources and Services Administration, US Department of Health and Human Services, Washington, DC.

We thank John M. Johnson, PhD, for his critical review of the manuscript.

Corresponding author and reprints: Myung K. Park, MD, Department of Pediatrics, MSC 7734, University of Texas Health Science Center, 7703 Floyd Curl Dr, San Antonio, TX 78229-3900 (e-mail: parkm@uthscsa.edu).

Friesen  RHLichtor  JL Indirect measurement of blood pressure in neonates and infants utilizing an automatic noninvasive oscillometric monitor. Anesth Analg. 1981;60742- 745
Link to Article
Kimble  KJDarnall  RA  JrYelderman  MAriagno  RLReam  AK An automated oscillometric technique for estimating mean arterial pressure in critically ill newborns. Anestheiology. 1981;54423- 425
Link to Article
Yelderman  MReam  AK Indirect measurement of blood pressure in the anesthetized patients. Anesthesiology. 1979;50253- 256
Link to Article
Park  MKMenard  SW Accuracy of blood pressure measurement by the Dinamap Monitor in infants and children. Pediatrics. 1987;79907- 914
Borow  KMNewburger  JW Noninvasive estimation of central aortic pressure using the oscillometric method for analysis of systemic artery pulsatile blood flow. Am Heart J. 1982;103879- 886
Link to Article
Weaver  MGPark  MKLee  D-H Difference in blood pressure levels obtained by auscultatory and oscillometric methods. AJDC. 1990;144911- 914
Chapman  JMClark  VACoulson  AHBrowning  GG Problems of measurement in blood pressure surveys: inter-observer differences in blood pressure determinations. Am J Epidemiol. 1966;84483- 494
American Heart Association, Blood Pressure Awareness Program Instructor's Manual.  Austin, Tex Texas Affiliate Inc1986;
Frolich  EDLabarth  DRMaxwell  MHPerloff  DWeidman  WH Recommendations for human blood pressure determination by sphygmomanometers: report of a Special Task Force appointed by the Steering Committee, American Heart Association. Circulation. 1988;77501A- 514A
Not Available, Update on the 1987 task force report on high blood pressure in children: a Working Group Report from the National High Blood Pressure Education Program. Pediatrics. 1996;98649- 658
Bland  JMAltman  DG Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8135- 160
Link to Article
O'Brien  EPetrie  JLittler  W  et al.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens. 1990;8607- 619
Link to Article
Gillman  MWCook  NR Blood pressure measurement in childhood epidemiological studies. Circulation. 1995;921049- 1057
Link to Article
Whincup  PHBruce  NGCook  DGShaper  AG The Dinamap 1846X automated blood pressure recorder: comparison with the Hawksley random zero sphygmomanometrr under field conditions. J Epidemiol Community Health. 1992;46164- 169
Link to Article
O'Brien  EMee  FAtkins  NO'Malley  K Accuracy of the Dinamap portable monitor, model 8100 determined by the British Hypertension Society protocol. J Hypertens. 1993;11761- 763
Link to Article

Figures

Place holder to copy figure label and caption
Figure 1.

Magnitude of difference in systolic blood pressure readings by the Dinamap and auscultatory methods plotted against the means of systolic blood pressure readings by the 2 methods for boys aged 9 to 10 years. Two broken horizontal lines represent the 95% confidence intervals for the mean difference. The solid horizontal line represents the mean difference.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 2.

Comparison of blood pressure levels of fifth-grade students from 2 studies. Dinamap model 1846SX was used in the 1990 study6 and Dinamap model 8100 (Critikon, Tampa, Fla) in the current study. Although the populations were different, they are both fifth-grade students from the same school district. BP indicates blood pressure; DIN, Dinamap; AUS, auscultatory; and K5, Korotkoff phase V diastolic pressure.

Graphic Jump Location

Tables

Table Graphic Jump LocationTable 1. Auscultatory and Dinamap Blood Pressure Readings According to the Order of Measurements*
Table Graphic Jump LocationTable 2. Mean Difference Between Dinamap* and Auscultatory Blood Pressure Readings According to Age Group†
Table Graphic Jump LocationTable 3. Mean Difference in Dinamap* and Auscultatory Blood Pressure Readings by Ethnicity and Sex†

References

Friesen  RHLichtor  JL Indirect measurement of blood pressure in neonates and infants utilizing an automatic noninvasive oscillometric monitor. Anesth Analg. 1981;60742- 745
Link to Article
Kimble  KJDarnall  RA  JrYelderman  MAriagno  RLReam  AK An automated oscillometric technique for estimating mean arterial pressure in critically ill newborns. Anestheiology. 1981;54423- 425
Link to Article
Yelderman  MReam  AK Indirect measurement of blood pressure in the anesthetized patients. Anesthesiology. 1979;50253- 256
Link to Article
Park  MKMenard  SW Accuracy of blood pressure measurement by the Dinamap Monitor in infants and children. Pediatrics. 1987;79907- 914
Borow  KMNewburger  JW Noninvasive estimation of central aortic pressure using the oscillometric method for analysis of systemic artery pulsatile blood flow. Am Heart J. 1982;103879- 886
Link to Article
Weaver  MGPark  MKLee  D-H Difference in blood pressure levels obtained by auscultatory and oscillometric methods. AJDC. 1990;144911- 914
Chapman  JMClark  VACoulson  AHBrowning  GG Problems of measurement in blood pressure surveys: inter-observer differences in blood pressure determinations. Am J Epidemiol. 1966;84483- 494
American Heart Association, Blood Pressure Awareness Program Instructor's Manual.  Austin, Tex Texas Affiliate Inc1986;
Frolich  EDLabarth  DRMaxwell  MHPerloff  DWeidman  WH Recommendations for human blood pressure determination by sphygmomanometers: report of a Special Task Force appointed by the Steering Committee, American Heart Association. Circulation. 1988;77501A- 514A
Not Available, Update on the 1987 task force report on high blood pressure in children: a Working Group Report from the National High Blood Pressure Education Program. Pediatrics. 1996;98649- 658
Bland  JMAltman  DG Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8135- 160
Link to Article
O'Brien  EPetrie  JLittler  W  et al.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens. 1990;8607- 619
Link to Article
Gillman  MWCook  NR Blood pressure measurement in childhood epidemiological studies. Circulation. 1995;921049- 1057
Link to Article
Whincup  PHBruce  NGCook  DGShaper  AG The Dinamap 1846X automated blood pressure recorder: comparison with the Hawksley random zero sphygmomanometrr under field conditions. J Epidemiol Community Health. 1992;46164- 169
Link to Article
O'Brien  EMee  FAtkins  NO'Malley  K Accuracy of the Dinamap portable monitor, model 8100 determined by the British Hypertension Society protocol. J Hypertens. 1993;11761- 763
Link to Article

Correspondence

CME
Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.
Submit a Comment

Multimedia

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Collections
PubMed Articles